Availability: | |
---|---|
IRHD-MA
EBP INSTRUMENTS
Shanghai, Shenzhen
90 sets per month
TT in advance
1 set
Features:
1. Using laser spot finding technology, it is more suitable for O-ring detection. The equipment will automatically find the highest point of the cut surface of the O-ring and send this point under the probe to automatically complete the hardness test.
2. Using a touch-control integrated operating system, the interface is intuitive and generous, and easier to operate
3. The servo system imported from Germany ensures precise positioning to ensure accurate measurement.
4. It comes with a micro-printer, and each sample supports 5 points of output.
Specification:
Y Axis maximum testing range: 30mm; Displacement accuracy: 0.001mm
X Axis maximum testing range: 120mm; Displacement accuracy: 0.001mm
O-rings minimum diameter: 0.8mm;
Flat test piece minimum thickness: 1mm (less than 1mm can layup)
Test Accuracy: 0.1IRHD
Test |
Diameters mm | Force on Ball | Force on Foot mN |
Test Range (IRHD) | Dwell Time (s) | ||
Contact mN | Indenting mN | Total mN | |||||
Method M (Microtest) | Ball: 0.395±0.005 Foot: 3.35±0.15 Hole: 1.00±0.15 |
8.3±0.5
|
145±0.5
| 153.3±1.0 |
235±30
| 30-85 | 30 |
Features:
1. Using laser spot finding technology, it is more suitable for O-ring detection. The equipment will automatically find the highest point of the cut surface of the O-ring and send this point under the probe to automatically complete the hardness test.
2. Using a touch-control integrated operating system, the interface is intuitive and generous, and easier to operate
3. The servo system imported from Germany ensures precise positioning to ensure accurate measurement.
4. It comes with a micro-printer, and each sample supports 5 points of output.
Specification:
Y Axis maximum testing range: 30mm; Displacement accuracy: 0.001mm
X Axis maximum testing range: 120mm; Displacement accuracy: 0.001mm
O-rings minimum diameter: 0.8mm;
Flat test piece minimum thickness: 1mm (less than 1mm can layup)
Test Accuracy: 0.1IRHD
Test |
Diameters mm | Force on Ball | Force on Foot mN |
Test Range (IRHD) | Dwell Time (s) | ||
Contact mN | Indenting mN | Total mN | |||||
Method M (Microtest) | Ball: 0.395±0.005 Foot: 3.35±0.15 Hole: 1.00±0.15 |
8.3±0.5
|
145±0.5
| 153.3±1.0 |
235±30
| 30-85 | 30 |